Installation
The Illumina MiSeqDx Integration Package v1.10.0 supports MiSeqDx instruments running in Diagnostic mode. For MiSeqDx instruments running in Research mode, install the MiSeq Integration Package. For more information, refer to MiSeq Integration v8.2.0 Installation.
Compatibility
MiSeqDx Integration v1.10.0 is compatible with the following software:
Clarity LIMS v6.2 and later
Secret Util v1.0 and later
IPP v2.6 and later
Prerequisites
MiSeqDx Integration v1.10.0 has the following prerequisites:
Mount run data network-attached storage (NAS) share
IPP is installed
Installation
MiSeqDx Integration v1.10.0 supports both on-premise and cloud integrations. This integration is distributed as the following RPM packages:
BaseSpaceLIMS-miseqdx-extensions
BaseSpaceLIMS-miseqdx-sequencing-service
The BaseSpaceLIMS-miseqdx-extensions RPM installs the following items:
Protocols and workflows
Database properties that configure the service
Placement pattern files that determine reagent index assignment
miseqdx-extensions.jar
miseqdx-sequencing-report.jar
The BaseSpaceLIMS-miseqdx-sequencing-service RPM installs the following items:
If not found, user configuration (the glsjboss user and the glsjdk8 and claritylims user groups)
If not installed, Java 8
Bash scripts to run the miseqdx_seqservice
miseqdx-sequencing.jar
The gls_events_mos_rta.bat batch file that configures RTA
Smoke test directories
On-Premise Installation
Use the following instructions to install the BaseSpaceLIMS-miseqdx-extensions and BaseSpaceLIMS-miseqdx-sequencing-service RPMs on the Clarity LIMS server.
Cloud Installation
The BaseSpaceLIMS-miseqdx-extensions RPM must be installed on the Clarity LIMS server. The BaseSpaceLIMS-miseqdx-sequencing-service RPM can be installed remotely on another server within the network.
Workflows, Protocols, and Steps Installed
MiSeqDx Integration v1.10.0 installs the following protocols:
CF 139-Variant Assay Library Prep 1.2
CF Clinical Sequencing Assay Library Prep 1.2
Universal Kit Library Prep 1.2
Illumina SBS MiSeqDx (CF 139-Variant Assay) 1.2
Illumina SBS MiSeqDx (CF Clinical Sequencing Assay) 1.2
Illumina SBS MiSeqDx (Universal Kit) 1.2
The integrations also install the following validation protocols that are included in a workflow with the same name:
MiSeqDx Validation (CF 139-Variant Assay) 1.2
MiSeqDx Validation (CF Clinical Sequencing Assay) 1.2
MiSeqDx Validation (Universal Kit) 1.2
For descriptions of the protocol and the steps, refer to MiSeqDx Integration v1.10.0 Configuration. For instructions on user interaction for each step and using the MiSeqDx validation workflows to validate the automated sample sheet generation, refer to MiSeqDx Integration v1.10.0 User Interaction, Validation and Troubleshooting.
Instrument Software
The instrument software is divided into the following modules:
MiSeqDx Operating Software (MOS) — Controls the instrument operation, including various configuration settings. This software is installed and runs on the instrument.
MiSeqDx Reporting Software (MRS) — Performs the following secondary analysis functions:
Demultiplexing
Alignment
Variant calling
Report generation
The specific functions that are supported vary by the kit. This software is installed on or off the instrument.
Real-Time Analysis (RTA) — Performs image processing and base calling (primary analysis). The software makes sure that data files are created and copied to the final destination folder and is installed and runs on the instrument.
Illumina User Management (IUM) — Contains a user database file that is used with the MiSeqDx instrument. This file controls user passwords and privileges for MOS.
For more information on the MiSeqDx software, refer to the MiSeqDx documentation at support.illumina.com.
Instrument Integration
Illumina provides a supported mechanism for using custom scripts on key events during a sequencing run. The Clarity LIMS support team has created batch files that plugs into these events. When the batch file is used, it reads the event information and writes it in a TXT event file at the same network share location that the instrument uses to write the run data. Another process running on the server where the sequencing service RPM is installed receives the event files and takes the appropriate actions.
The sequencing service monitors the following events (the actual event names may be different):
End Run — This event is used to update the sequencing steps in Clarity LIMS, captures key process data and files, and parses run statistics for output custom fields.
Begin Secondary Analysis — Indicates that secondary analysis in the MRS has started so that the sequencing service can start to monitor for results. After secondary analysis is complete, the VCF files are uploaded to Clarity LIMS.
Validation
Last updated