Last updated
Was this helpful?
These Release Notes describe the key changes to software components for NovaSeq 6000 Integration Package v2.4.1.
This version of the integration package is compatible with the following software:
Clarity LIMS v5.1 and later
NGS Extensions Package v5.21 and later
NovaSeq Control Software (NVCS) v1.8.0
Illumina bcl2fastq v2.20
Features provided with this release of the integration package are as follows:
Updates to the Validate Flowcell Barcode automation (in the Load to Flowcell step of the NovaSeq Xp protocol) to support future flow cell barcode formats.
Fixed the Validate Flowcell Barcode automation (in the Load to Flowcell Step of NovaSeq Xp Protocol) to accept only the upper case alphanumeric barcode format.
Fixed the following Calculate Volumes automation script values to sync with the NovaSeq 6000 System Denature and Dilute Libraries Guide (document # 1000000106351):
PhiX
NaOH
Tris-HCl
Bulk pool volume
None
Last updated
Was this helpful?
Was this helpful?