NovaSeq 6000 File-based v2.4.1

These Release Notes describe the key changes to software components for NovaSeq 6000 Integration Package v2.4.1.

Compatibility

This version of the integration package is compatible with the following software:

  • Clarity LIMS v5.1 and later

  • NGS Extensions Package v5.21 and later

  • NovaSeq Control Software (NVCS) v1.8.0

  • Illumina bcl2fastq v2.20

New Features

Features provided with this release of the integration package are as follows:

  • Updates to the Validate Flowcell Barcode automation (in the Load to Flowcell step of the NovaSeq Xp protocol) to support future flow cell barcode formats.

Defects Fixed

  • Fixed the Validate Flowcell Barcode automation (in the Load to Flowcell Step of NovaSeq Xp Protocol) to accept only the upper case alphanumeric barcode format.

  • Fixed the following Calculate Volumes automation script values to sync with the NovaSeq 6000 System Denature and Dilute Libraries Guide (document # 1000000106351):

    • PhiX

    • NaOH

    • Tris-HCl

    • Bulk pool volume

Known Issues

None