iScan v1.2.0

The integration includes the following:

  • Automated beadchip accessioning in Clarity LIMS (via Illumina Connected Analytics during pre-imaging verification).

  • Automated verification of beadchip status. Samples are automatically created upon successful verification.

  • Automated beadchip imaging status tracking (imaging run started; errored out; completed successfully; or ended by user events).

  • Manual invocation of analysis on Illumina Connected Analytics.

  • Automated analysis status (for completed successfully or errored out events) tracking.

iScan

The Illumina iScan integration provides preconfigured workflows that map to established lab protocols and steps used with the iScan system. The integration also supports the associated consumables and assays used in these protocols.

iScan System

BeadChip Accessioning, Imaging, and Analysis

The Illumina iScan Integration Package v1.2.0 supports the integration of Clarity LIMS to iScan systems. The package includes the following workflows, which are included in the Illumina Preset Protocols (IPP) Package v2.9.0:

  • iLASS Infinium Genotyping Assay v1.1

  • iLASS Infinium Genotyping Assay with PGx v1.1

This integration enables you to do the following tasks:

  • Accession the BeadChip.

  • Handle the imaging run event (including pre-imaging verification for the imaging step).

  • Handle the analysis run event for the analysis step.

triangle-exclamation

The configuration used in this integration supports iScan lab processes. Any changes to the configuration (including renaming protocols, steps, and fields) can cause issues with these processes.

hashtag
Accession the BeadChip

Beadchip are auto accessioned for the workflows listed in Supported Workflows. The beadchip containers are created in Clarity by iLASS at the hybridization step of the post-lab protocol. Prior to imaging, a verification is performed and beadchips that are not accessioned prior to start of the workflows will be automatically accessioned by Illumina Connected Analytics (ICA). The BeadChip container type is defined based on the BeadChip product type (eg, PGx) and the product version. Because every BeadChip has a unique barcode, Clarity LIMS can track usage throughout the entire workflow. The following BeadChip names are supported in iScan Integration v1.2.0:

  • GCRA-24v1-0

  • GDA_PGx-8v1-0

  • GSA-48v4-0

  • MSA-48v1-0

hashtag
Image Infinium BeadChip on iScan Step

Master Step: Image Infinium BeadChip on iScan (iLASS Infinium Genotyping Assay v1.1) or Image Infinium BeadChip on iScan (iLASS Infinium Genotyping Assay with PGx v1.1)

triangle-exclamation

This step is automated and must not be started manually.

hashtag
Master Step Custom Fields

The following table lists field configuration details defined on the Image Infinium BeadChip on iScan master step. These fields are located under Configuration, on the Master Step Fields tab.

triangle-exclamation

Do not delete, rename, or modify. These actions will break the integration.

hashtag
Pre-Imaging Verification

When BeadChips are loaded onto the iScan system, the iScan Control Software (iCS) with LIMS mode enabled performs pre-imaging verification to make sure that user loads the correct BeadChip. This verification is done as follows.

  1. iScan Control Software verifies the BeadChip container is queued to the imaging step in Clarity LIMS.

  2. After the verification completes sucessfully, Clarity LIMS iScan integration creates run on ICA and starts BeadChip imaging starts automatically.

hashtag
Imaging Run Status Tracking

When the BeadChip is scanned with an iScan, Clarity LIMS receives the events of the run status update. Then, the custom fields are automatically populated with information.

The Run Status field can show the following statuses:

  • Running

  • Completed

  • Aborted

  • Failed

If there is a run error event, the Run Complete Time field is not populated.

hashtag
Invoke Infinium Analysis Step

Master Step: Invoke Infinium Analysis (iLASS Infinium Genotyping Assay v1.1) or Invoke Infinium Analysis (iLASS Infinium Genotyping Assay with PGx v1.1)

In this step, the iScan Integration handles events related to BeadChip analysis. This step is a manual step that requires data input of Analysis Configuration ID and Analysis Run Name fields to start the analysis run.

hashtag
Master Step Custom Fields

The following table lists field configuration details defined on the Invoke Infinium Analysis master step. These fields are located under Configuration, on the Master Step Fields tab.

hashtag
Configure and Start an Analysis Run

  1. On the Record Details milestone, navigate to Step Details and enter the Analysis Run Name. The Analysis Run Name can only contain a maximum of 255 alphanumeric characters, underscores, hyphens, and periods. The name cannot start with the wfr. characters.

  2. Enter the Analysis Configuration ID. To retrieve the configuration ID that is associated with the analysis, contact Illumina Support.

  3. Select Start Analysis v1.0 to start the analysis run.

hashtag
Analysis Run Status Tracking

When the analysis is started on ICA, Clarity LIMS receives the event and displays related information. The custom fields are automatically populated with information.

The Analysis Status field can show the following statuses:

  • Analysis Running

  • Analysis Completed

  • Analysis Completed with Warning

  • Analysis Aborted

If there is an analysis error event, the Analysis Complete Time field is not populated.

hashtag
Properties Installed

Refer to for the properties installed with Illumina iScan Integration Package v1.2.0.

Data Path

ℹ Automatically populated when imaging is started

Text

Read Only

Instrument Run ID

ℹ Automatically populated when imaging is started

Text

Read Only

Run Status

ℹ Automatically populated when imaging is started, completed, aborted, or has failed

Text

Read Only

DMAP Location

ℹ Automatically populated when imaging is started

Text

Read Only

Instrument Type

ℹ Automatically populated when imaging is started

Text

Read Only

Failure Reason

ℹ Automatically populated when imaging fails

Text

Read Only

Log

ℹ Automatically populated when imaging is started, completed, aborted, or has failed

Multiline Text

Read Only

ICA Run ID

ℹ Automatically populated when imaging is started

Text

Read Only

Run Complete Time

ℹ Automatically populated when imaging is completed or aborted

Text

Read Only

Comment

Multiline Text

Analysis Start Time

ℹ Automatically populated when analysis is started

Text

Read Only

Analysis Report

ℹ Automatically populated when analysis is completed

Text

Read Only

Analysis Status

ℹ Automatically populated when analysis is started, completed, aborted, or has failed

Text

Read Only

Analysis Run ID

ℹ Automatically populated when analysis is started

Text

Read Only

Log

ℹ Automatically populated when analysis is started, completed, aborted, or has failed

Multiline Text

Read Only

  • Analysis Timedout

  • Analysis Errored

  • Field Name

    Field Type

    Options

    Additional Options and Dropdown Items

    BeadChip Barcode

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Instrument

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Run Start Time

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Field Name

    Field Type

    Options

    Additional Options and Dropdown Items

    Analysis Complete Time

    ℹ Automatically populated when analysis is completed or aborted

    Text

    Read Only

    Analysis Run Name

    Text

    Required

    Analysis Configuration ID

    Text

    Required

    Integration Properties Details

    iScan v1.1.0

    The integration includes the following:

    • Automated beadchip accessioning in Clarity LIMS upon beadchip creation on Illumina Connected Analytics (ICA).

    • Automated verification of beadchip status. Samples are automatically created upon successful verification.

    • Automated beadchip imaging status tracking (imaging run started; errored out; completed successfully; or ended by user events).

    • Manual invocation of analysis on ICA.

    • Automated analysis status (for completed successfully or errored out events) tracking.

    The integration utilizes the following pre-configured workflows in IPP:

    • iLASS Infinium Genotyping Assay v1.0

    BeadChip Accessioning, Imaging, and Analysis

    The Illumina iScan Integration Package v1.1.0 supports the integration of Clarity LIMS to iScan systems. The package includes the following workflows, which are included in the Illumina Preset Protocols (IPP) Package v2.6.0:

    • iLASS Infinium Genotyping Assay v1.0

    • iLASS Infinium Genotyping Assay with PGx v1.0

    This integration enables you to do the following tasks:

    • Accession the BeadChip.

    • Handle the imaging run event (including pre-imaging verification for the imaging step).

    • Handle the analysis run event for the analysis step.

    triangle-exclamation

    The configuration used in this integration supports iScan lab processes. Any changes to the configuration (including renaming protocols, steps, and fields) can cause issues with these processes.

    hashtag
    Accession the BeadChip

    In the iScan workflows, BeadChips are accessioned into Clarity LIMS before the hybridization step of the post-lab protocol. The BeadChip container type is defined based on the BeadChip product type (eg, PGx) and the product version. Because every BeadChip has a unique barcode, Clarity LIMS can track usage throughout the entire workflow. The following BeadChip names are supported in iScan Integration v1.1.0:

    • GCRA-24v1-0

    • GDA_PGx-8v1-0

    • GSA-48v4-0

    If you need help accessioning your BeadChip, contact Illumina Support.

    hashtag
    Image Infinium BeadChip on iScan Step

    Master Step: Image Infinium BeadChip on iScan (iLASS Infinium Genotyping Assay v1.0) or Image Infinium BeadChip on iScan (iLASS Infinium Genotyping Assay with PGx v1.0)

    triangle-exclamation

    This step is automated and must not be started manually.

    hashtag
    Master Step Custom Fields

    The following table lists field configuration details defined on the Image Infinium BeadChip on iScan (iLASS Infinium Genotype Assay/with PGx v1.0) master step. These fields are located under Configuration, on the Master Step Fields tab.

    triangle-exclamation

    Do not delete, rename, or modify. These actions will break the integration.

    hashtag
    Pre-Imaging Verification

    When BeadChips are loaded onto the iScan system, the iScan Control Software (iCS) with LIMS mode enabled performs pre-imaging verification to make sure that user loads the correct BeadChip. This verification is done as follows.

    1. iScan Control Software verifies the BeadChip container is queued to the imaging step in Clarity LIMS.

    2. After the verification completes sucessfully, Clarity LIMS iScan integration creates run on ICA and starts BeadChip imaging starts automatically.

    hashtag
    Imaging Run Status Tracking

    When the BeadChip is scanned with an iScan, Clarity LIMS receives the events of the run status update. Then, the custom fields are automatically populated with information.

    The Run Status field can show the following statuses:

    • Running

    • Completed

    • Aborted

    • Failed

    If there is a run error event, the Run Complete Time field is not populated. The Failure Reason field will be populated only when the run error event happens.

    hashtag
    Invoke Infinium Analysis Step

    Master Step: Invoke Infinium Analysis (iLASS Infinium Genotyping Assay v1.0) or Invoke Infinium Analysis (iLASS Infinium Genotyping Assay with PGx v1.0)

    In this step, the iScan Integration handles events related to BeadChip analysis. This step is a manual step that requires data input of Analysis Configuration ID and Analysis Run Name fields to start the analysis run.

    hashtag
    Master Step Custom Fields

    The following table lists field configuration details defined on the Invoke Infinium Analysis (iLASS Infinium Genotyping Assay/with PGx v1.0) step. These fields are located under Configuration, on the Master Step Fields tab.

    hashtag
    Configure and Start an Analysis Run

    1. On the Record Details milestone, navigate to Step Details and enter the Analysis Run Name. The Analysis Run Name can only contain a maximum of 255 alphanumeric characters, underscores, hyphens, and periods. The name cannot start with the wfr. characters.

    2. Enter the Analysis Configuration ID. To retrieve the configuration ID that is associated with the analysis, contact Illumina Support.

    3. Select Start Analysis v1.0 to start the analysis run.

    hashtag
    Analysis Run Status Tracking

    When the analysis is started on ICA, Clarity LIMS receives the event and displays related information. The custom fields are automatically populated with information.

    The Analysis Status field can show the following statuses:

    • Analysis Running

    • Analysis Completed

    • Analysis Completed with Warning

    • Analysis Aborted

    If there is an analysis error event, the Analysis Complete Time field is not populated.

    Data Path

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Instrument Run ID

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Run Status

    ℹ Automatically populated when imaging is started, completed, aborted, or has failed

    Text

    Read Only

    DMAP Location

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Instrument Type

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Failure Reason

    ℹ Automatically populated when imaging fails

    Text

    Read Only

    Log

    ℹ Automatically populated when imaging is started, completed, aborted, or has failed

    Multiline Text

    Read Only

    ICA Run ID

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Run Complete Time

    ℹ Automatically populated when imaging is completed or aborted

    Text

    Read Only

    Analysis Start Time

    ℹ Automatically populated when analysis is started

    Text

    Read Only

    Analysis Report

    ℹ Automatically populated when analysis is completed

    Text

    Read Only

    Analysis Status

    ℹ Automatically populated when analysis is started, completed, aborted, or has failed

    Text

    Read Only

    Analysis Run ID

    ℹ Automatically populated when analysis is started

    Text

    Read Only

    Log

    ℹ Automatically populated when analysis is started, completed, aborted, or has failed

    Multiline Text

    Read Only

  • Analysis Timedout

  • Analysis Errored

  • Field Name

    Field Type

    Options

    Additional Options and Dropdown Items

    BeadChip Barcode

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Instrument

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Run Start Time

    ℹ Automatically populated when imaging is started

    Text

    Read Only

    Field Name

    Field Type

    Options

    Additional Options and Dropdown Items

    Analysis Complete Time

    ℹ Automatically populated when analysis is completed or aborted

    Text

    Read Only

    Analysis Run Name

    Text

    Required

    Analysis Configuration ID

    Text

    Required

    iScan System v1.0

    hashtag
    Overview

    The iScan workflow includes the following functionality:

    • Preset protocol.

    • Ability to attach the scan log to the samples.

    hashtag
    Protocol 1: Infinium Array Imaging (iScan v1.0)

    Protocol Type = Other

    Next Steps Configuration

    hashtag
    Step 1: Image Infinium BeadChip (iScan v1.0)

    • Master Step Name = Image Infinium BeadChip - iScan v1.0

    • Step Type = No Outputs

    hashtag
    Automations

    chevron-rightSet Array Secondary Analysishashtag
    • Trigger Location = Record Details

    • Trigger Style = Automatic upon entry

    chevron-rightSet Next Step - Advancehashtag
    • Trigger Location = Record Details

    • Trigger Style = Automatic upon exit

    chevron-rightRouting Script - GenomeStudiohashtag
    • Trigger Location = Step

    • Trigger Style = Automatic upon exit

    hashtag
    Queue/Ice Bucket

    • Defaults

      • Sample Grouping = Group by Containers

      • Well Sort Order = Row

    hashtag
    Record Details

    • Step Data (Master Step Fields)

    ℹ The workflow version and step version for the routing script may vary depending on the version of the IPP.
    Sample Table (Column Headers)

    Category

    Field Name

    Field Type

    Options

    Additional Options and Dropdown Items

    Container

    Container Name

    Built-in

    Container

    LIMS ID (Container)

    Built-in

    Step File Placeholders
    • Log - Automatically attached

    • Scan Log - Manually uploaded

  • Sample Table

    • Sample Display Default = Collapse

    • Well Sort Order = Row

    • Table Columns - Global Fields

  • Field Name

    Field Type

    Options

    Additional Options and Dropdown Items

    Comment

    Multiline Text

    bash -l -c "/opt/gls/clarity/bin/java -jar /opt/gls/clarity/extensions/ngs-common/v5/EPP/ngs-extensions.jar -i {stepURI:v2} -u {username} -p {password} script:evaluateDynamicExpression -exp 'if (input.::Infinium Kit Name:: == ::Infinium MethylationEPIC Kit:: ) { input.::Array Secondary Analysis:: = ::GS-METHYLATION::} ; if (input.::Infinium Kit Name:: != ::Infinium MethylationEPIC Kit:: ) { input.::Array Secondary Analysis:: = ::GS-GENOTYPING::}' -log {compoundOutputFileLuid0}"
    bash -l -c "/opt/gls/clarity/bin/java -jar /opt/gls/clarity/extensions/ngs-common/v5/EPP/ngs-extensions.jar -i {stepURI:v2} -u {username} -p {password} \
          script:evaluateDynamicExpression \
          -t false \
          -h false \
          -exp 'nextStep = ::ADVANCE::' \
          -log {compoundOutputFileLuid0}"
    bash -c "/opt/gls/clarity/bin/java -jar /opt/gls/clarity/extensions/ngs-common/v5/EPP/ngs-extensions.jar -u {username} -p {password} -i {stepURI:v2} -l {compoundOutputFileLuid0} script:changeWorkflow \
    \
    --FIELD_NAME 'Array Secondary Analysis' \
    --FIELD_VALUE 'GS-GENOTYPING' \
    --WORKFLOW 'GenomeStudio v1.0' \
    --STEP 'Create Infinium Genotyping Analysis Sample Sheet (GenomeStudio v1.0)' \
    --INPUTS_OR_OUTPUTS 'INPUTS' \
    --FIELD_NAME 'Array Secondary Analysis' \
    --FIELD_VALUE 'GS-METHYLATION' \
    --WORKFLOW 'GenomeStudio v1.0' \
    --STEP 'Create Infinium Methylation Analysis Sample Sheet (GenomeStudio v1.0)' \
    --INPUTS_OR_OUTPUTS 'INPUTS'"

    BeadChip Position

    Text

    Read Only

    Derived Sample

    DNA Plate Barcode

    Text

    Read Only

    Derived Sample

    Infinium Kit Name

    Text

    Read Only

    Derived Sample

    Sample Name

    Built-in

    Derived Sample

    Sample_Well

    Text

    Project

    Project Name

    Built-in

    Container

    Well

    Built-in

    Derived Sample

    Sample Name

    Built-in

    Derived Sample

    Waiting

    Built-in

    Project

    Project Name

    Built-in

    Category

    Field Name

    Field Type

    Options

    Additional Options and Dropdown Items

    Derived Sample

    Array Secondary Analysis

    Text Dropdown

    • Presets

      • GS-GENOTYPING

      • GS-METHYLATION

    • Default = GS-GENOTYPING

    Derived Sample

    BeadChip Barcode

    Text

    Read Only

    Derived Sample

    Release Notes

    Last Updated: November 2024

    Release Date: July 2024

    Document Version: 2

    These release notes describe the key changes to software components for the Clarity LIMS iScan Integration Package version 1.2.0.

    hashtag
    Compatibility

    Refer to under Instruments & Integrations.

    hashtag
    New Features

    • Updated iScan integration to support 'ILASS Infinium Genotyping Assay v1.1' and 'ILASS Infinium Genotyping Assay with PGx v1.1' workflows

    • Supports the new 'Infinium Methylation Array EX-48' beadchip

    • Removed unused Analysis Result column from Sample Details table in Infinium Analysis step.

    hashtag
    Defects fixed

    • Fixed imaging result not reflecting correctly in the "Infinium Array Imaging on iScan" step upon re-imaging of beadchip.

    • Updated iScan install details to table retrieved by getInstallHistory script.

    hashtag
    Known issues

    • Warning log message displays when the ImagingAborted and ImagingFailed events are processed.

    hashtag
    Revision History

    Integration-related properties can now be accessed and updated via System Setting in Clarity v6.3. Refer to for configurable properties.
  • Updated integration properties to be extensible to future array projects

    • Replaced supportedImagingWorkflowStepNames with supportedImagingWorkflowStepNamePattern

    • Replaced supportedAnalysisWorkflowStepNames with supportedAnalysisWorkflowStepNamePattern

  • Added analysis run ID in analysis run logging to improve troubleshooting.

  • Version

    Changes

    2

    • Updated Compatibility section to reference Compatibility matrix table.

    1

    • Initial release.

    Compatibility
    Integration Properties Details

    Release Notes

    Last Updated: November 2024

    Release Date: September 2023

    Document Version: 2

    These release notes describe the key changes to software components for the Clarity LIMS iScan Integration Package version 1.1.0.

    hashtag
    Compatibility

    Refer to under Instruments & Integrations.

    hashtag
    New Features

    • Updates to the Java and third-party dependency libraries used in the integration.

    hashtag
    Defects Fixed

    • Fixed the SIS core service not being properly registered in systemctl.

    hashtag
    Known Issues

    • Warning log message displays when the ImagingAborted and ImagingFailed events are processed.

    hashtag
    Revision History

    Version

    Changes

    2

    • Updated Compatibility section to reference Compatibility matrix table.

    1

    • Initial release.

    Compatibility