Manual Upgrade

The NextSeq 500/550 Integration v2.5.0 integration package requires the NextSeq 500/550 v1.3 workflow available from IPP v2.10. If you do not want to upgrade through the IPP, you can manually upgrade the workflow configuration from NextSeq 500/550 v1.2 workflow to v1.3 workflow by following instructions in this guide to generate samplesheet compatible with LRM v4.0.

Generate NextSeq 500/550 Samplesheet Compatible with LRM v4.0

To support the generation of LRM v4.0-compatible samplesheet, you must modify the Denature and Dilute step, and replacement of template files.

Modify Denature and Dilute Step

Automation

  1. From Configuration, select the Automation tab.

  2. Rename the Generate bcl2fastq2 NextSeq Samplesheet automation to Generate NextSeq Samplesheet.

The default command line for the automation remains unchanged.

Master Step Fields

  1. From Configuration, select the Custom Fields tab.

  2. Select the Denature and Dilute Master Step.

  3. Add the following master step fields:

    Field Name

    Field Type

    Field Constraints/Options

    Preset Values/Additional Options and Drop-Down Items

    Comment

    Multiline Text

    Genome Folder

    Text

    Manifest

    Text

    Samplesheet Template

    Text Dropdown

    Required Field

    Presets

    • NextSeq_Samplesheet.csv

    • NextSeq_ReverseComplement_Samplesheet.csv

    Taxonomy Database

    Text

  4. Remove the following master step fields:

    • Adapter

    • Adapter Read 2

    • Mask Adapter

    • Mask Adapter Read 2

    • SampleSheet Template

  5. Update the following master step fields:

    Field Name

    Changes Required

    Screenshot

    Experiment Name

    • Update Required Field to Yes

    Workflow

    • Update Dropdown presets to

      • Assembly

      • DNA Amplicon

      • DNA Enrichment

      • GenerateFASTQ

      • Library QC

      • 16S Metagenomics

      • PCR Amplicon

      • Resequencing

      • RNA Fusion

      • Small RNA

Group of Defaults

  1. From Configuration, select the Custom Fields tab.

  2. Select the Denature and Dilute Master Step.

  3. Add the following Group of Defaults:

    Group of Defaults

    Details

    Screenshot

    RNA Fusion

    • Read 1 Cycles = 251

    • Workflow = RNA Fusion

    Small RNA

    • Read 1 Cycles = 251

    • Workflow = Small RNA

  4. Update the following Group of Defaults:

    Group of Defaults

    Changes Required

    Screenshot

    Custom Amplicon

    • Rename to DNA Amplicon

    • Update Workflow field to DNA Amplicon

    Enrichment

    • Rename to DNA Enrichment

    • Update Workflow field to DNA Enrichment

    Generate FASTQ

    • Rename to GenerateFASTQ

    Library QC

    • Update Workflow field to Library QC

    Metagenomics

    • Rename to 16S Metagenomics

    • Update Workflow field to 16S Metagenomics

Step File Placeholders

  1. From Configuration, select the Lab Work tab.

  2. Select the Denature and Dilute Master Step.

  3. Rename the following Step File Placeholders under Record Details milestone:

    • from bcl2fastq SampleSheet to Samplesheet

    • from bcl2fastq SampleSheet Generation Log to Samplesheet Generation Log

Replace Template Files

Replace the following template files at /opt/gls/clarity/extensions/conf/driverfiletemplates/:

  • BCL2FASTQ_Reverse_Complement_Samplesheet.csv with NextSeq_ReverseComplement_Samplesheet.csv

  • BCL2FASTQ_Samplesheet.csv with NextSeq_Samplesheet.csv

Contact Illumina Support to replace the files if required.