Configuration
The Illumina MiSeq Integration Package v8.3.0 supports the integration of Clarity LIMS to Illumina MiSeq Sequencing Systems.
The integration allows for automated tracking of an Illumina sequencing run in Clarity LIMS. This capability includes tracking sequencing run status, generating run report, and capturing and parsing run statistics. In addition, this integration provides automated generation of a sample sheet file for use with the MiSeq Control Software (MCS) and Local Run Manager (LRM).
For instructions on user interaction for each step, validating and troubleshooting the Illumina MiSeq Integration Package, refer to MiSeq Integration v8.3.0 User Interaction, Validation and Troubleshooting.
Compatibility
The MiSeq Sequencing v3.2 workflow is compatible with MiSeq Integration Package v8.2.0 and v8.3.0.
MiSeq Integration v8.2.0 and above is only required for MiSeq Control Software (MCS) v4.0. For MCS v3.1 or earlier, do not upgrade to MiSeq Integration v8.2.0 or above. This upgrade breaks the integration.
Prerequisites and Assumptions
Before samples are assigned to the MiSeq Sequencing v3.2 workflow, make sure that the following prerequisites are completed:
Samples have been accessioned into Clarity LIMS.
Samples have been run through QC and library prep.
Samples have been normalized, and the value is captured in a field called Normalized Molarity (nM).
For more information on sample accessioning, refer to Sample Accessioning and Upload and Modify Samples in the Clarity LIMS (Clarity & LabLink Reference Guide) documentation.
Samples can be assigned to the MiSeq Sequencing v3.2 workflow automatically using a routing script or manually from the Projects & Samples dashboard. Refer to Assign and Process Samples in the Clarity LIMS (Clarity & LabLink Reference Guide) documentation.
Workflows, Protocols, and Steps
The Illumina MiSeq Integration includes the MiSeq Sequencing v3.2 workflow, which contains a single protocol of the same name.
Step 1: Library Pooling (MiSeq v3.2)
In this step, the lab scientist manually places libraries into pools in the Clarity LIMS Placement screen.
Master Step Fields
The following field is configured on the Library Pooling (MiSeq v3.2) master step. The field displays on the Record Details screen at run time.
Library Pooling (MiSeq v3.2) Master Step Field Configuration
Field Name | Field Type | Field Constraints/Options | Preset Values/Additional Options and Drop-down Items |
Comment | Multiline Text |
Global Fields
The following table lists the global fields that are displayed on the Queue and Ice Bucket screens of the Library Pooling (MiSeq v3.2) step. Most fields display in expanded view only.
Global Field Configuration (Submitted Sample)
Field Name | Field Type | Field Constraints/Options | Preset Values/Additional Options and Drop-Down Items |
Application | Text Dropdown | Custom Entries | Presets
|
Pooling | Text Dropdown | Custom Entries | Presets
|
Read Length | Text | ||
Sequencing Coverage | Text | ||
Sequencing Method | Text Dropdown | Custom Entries | Presets
|
Global Field Configuration (Derived Sample)
Field Name | Field Type | Field Constraints/Options | Preset Values/Additional Options and Drop-Down Items |
Normalized Molarity (nM) | Numeric | Decimal places displayed = 2 |
Step 2: Denature and Dilute (MiSeq v3.2)
In this step, pooled libraries are denatured and diluted, and then placed into the reagent cartridge loaded into the MiSeq instrument.
Master Step Fields
Most fields configured on the Denature and Dilute (MiSeq v3.2) step display on the Record Details screen in the Step Data table.
These fields are manually populated at run time. The values are then used to generate the sample sheet.
Denature & Dilute (MiSeq v3.2) Master Step Field Configuration
Field Name | Field Type | Options | Additional Options and Drop-Down Items |
Adapter Read 1 | Text | ||
Adapter Read 2 | Text | ||
Aligner | Text Dropdown |
| |
Aligner-Key | Text |
|
|
Aligner-Value | Text |
| |
Annotation | Text Dropdown |
| |
Applicable analysis fields for the selected Workflow | Multiline Text |
| |
Comment | Multiline Text | ||
Custom Primers | Text Dropdown |
| |
Description | Text | ||
Experiment Name | Text |
| |
Export to gVCF | Text Dropdown |
| |
Flag PCR Duplicates | Text Dropdown |
| |
Genome Folder | Text | ||
Indel Realignment | Text Dropdown |
| |
Indel Repeat Filter Cutoff | Text Dropdown |
| |
Indel-Realignment-Key | Text |
|
|
Indel-Realignment-Value | Text |
|
|
Manifest | Text | ||
Manifest Padding | Text Dropdown |
| |
Manifest-Section | Text |
|
|
Picard HS Metric Reporting | Text Dropdown |
| |
Read 1 Cycles | Numeric Dropdown |
|
|
Read 2 Cycles | Numeric Dropdown |
|
|
Read Stitching | Text Dropdown |
| |
Reverse Complement | Text Dropdown |
| |
UMI - Read 1 Length | Numeric |
| |
UMI - Read 1 Start From Cycle | Numeric |
| |
UMI - Read 2 Length | Numeric |
| |
UMI - Read 2 Start From Cycle | Numeric |
| |
Validation Script 1 | Multiline Text |
|
|
Validation Script 2 | Multiline Text |
|
|
Variant Caller | Text Dropdown |
| |
Variant Caller Depth Filter | Numeric |
| |
Variant Quality Filter | Numeric |
| |
Variant Frequency Percentage | Numeric |
| |
Variant-Caller-Value | Text |
|
|
Workflow | Text Dropdown |
|
|
Groups of Defaults
Global Fields
The following table shows the global fields that are configured to display on the Queue, Ice Bucket, and Record Details screens of the Denature and Dilute (MiSeq v3.2) step:
ℹ️ The Submitted Sample field, Progress, (added from previous MiSeq workflow) is obsolete in MiSeq v3.2.
Global Field Configuration (Submitted Sample)
Field Name | Field Type | Field Constraints/Options | Preset Values/Additional Options and Drop-Down Items |
Read Length | Text | ||
Sequencing Method | Text Dropdown |
|
|
Global Field Configuration (Derived Sample)
Field Name | Field Type | Field Constraints/Options | Preset Values/Additional Options and Drop-Down Items |
Final Loading Concentration | Numeric Dropdown |
|
|
Step File Placeholders
Placeholders for the following files are configured on the Record Details screen of the Denature and Dilute (MiSeq v3.2) step.
Step 3: MiSeq Run (MiSeq v3.2)
Master Step Fields
The following fields are configured on the MiSeq Run (MiSeq v3.2) step. These fields display on the Record Details screen at run time. The read-only fields are automatically populated at the end of the run.
MiSeq Run (MiSeq v3.2) Master Step Field Configuration
Field Name | Field Type | Options | Additional Options and Drop-Down Items |
Chemistry | Text | Read Only | |
Comment | Multiline Text | ||
Experiment Name | Text | Read Only | |
Finish Date | Date | Read Only | |
Flow Cell ID | Text | Read Only | |
Flow Cell Version | Text | Read Only | |
Index 1 Read Cycles | Numeric | Read Only | Decimal Places Displayed: 0 |
Index 2 Read Cycles | Numeric | Read Only | Decimal Places Displayed: 0 |
Output Folder | Text | Read Only | |
PR2 Bottle ID | Text | Read Only | |
Read 1 Cycles | Numeric | Read Only | Decimal Places Displayed: 0 |
Read 2 Cycles | Numeric | Read Only | Decimal Places Displayed: 0 |
Reagent Cartridge ID | Text | Read Only | |
Reagent Cartridge Part # | Text | Read Only | |
Run ID | Text | Read Only | |
Run Type | Text | Read Only | |
Status | Text | Read Only | |
Workflow | Text | Read Only |
Global Fields
There are several sample and measurement global fields that are displayed on the Record Details screen of the MiSeq Run (MiSeq v3.2) step. These fields are autopopulated at the end of the sequencing run.
For more information, refer to Sequencing Results Parsing.
Step File Placeholders
Placeholders for the following files are configured on the Record Details screen of the MiSeq Run (MiSeq v3.2) step:
Illumina Run Report (automatically attached)
Link to Run Folder (automatically attached)
Run Parameters (automatically attached)
Run Info (automatically attached)
Lab Tracking Form (manually uploaded)
Log File (automatically attached)
For details, refer to Generated and Captured Files.
Sample Sheet Generation
Sample sheet generation occurs in the Denature & Dilute (MiSeq v3.2) step. This step places samples on the container loaded in the system.
The default configuration provides only the Validate Run Setup and Generate MiSeq SampleSheet automation. This automation uses the Template File Generator (DriverFileGenerator.jar) and a template file to generate a CSV format file for use with the MiSeq Control Software (MCS).
The sample sheet content is determined by the fields on the Record Details screen of the step in the Step Data table. The values entered into these fields are used to populate the sample sheet.
To customize the template used to create the sample sheet, you can insert additional columns.
The following additional details are available:
For a sample template that you can download and customize for the lab, refer to Illumina Instrument Sample Sheets (NGS v5.17 & later).
For instructions on customizing the template, refer to Creating Template Files.
Sequencing Results Parsing
The MiSeq Run (MiSeq v3.2) step records information for the flow cell lane and generates a report summarizing the results. In addition, run parameters, run info, and a link to the run folder are automatically captured.
Generated and Captured Files
The following table describes the run information files, reports, placeholders, and links that Clarity LIMS automatically generates or captures during a sequencing run:
Run Information Generated or Captured by MiSeq Run (MiSeq v3.2) Step
Item | Description |
Run Info Run Parameters |
|
Link to Run Folder |
|
Illumina Run Report |
|
Lab Tracking Form |
|
Metadata
The following list shows metadata that Clarity LIMS automatically captures from the Illumina sequencing software as part of a sequencing run. This information is gathered from various run result files and events.
Chemistry
Experiment Name (entered in software)
Finish Date (run completion date)
If the End Run event contains a date in the format YYYY-MM-DD, Finish Date is set to the date in the event file.
If the End Run event does not contain a date or the date is in the wrong format, Finish Date is set to the date when the event file is processed.
Flow Cell ID
Flow Cell Version
Index 1 Read Cycles (intended Index cycles)
Index 2 Read Cycles (intended Index cycles)
Output Folder (run folder root)
PR2 Bottle ID
Reagent Cartridge ID
Reagent Cartridge Part #
Read 1 Cycles
Read 2 Cycles
Run ID (the unique run ID)
Run Type
Status (current status of the sequencing run on the instrument)
Workflow
Primary Analysis Metrics
The following table lists the real-time analysis (RTA) primary analysis metrics Clarity LIMS automatically captures and records per read, for samples in each flow cell lane. These metrics are captured upon run completion and are stored as fields in the Sample Details table on the Record Details screen.
To see both per read and per lane metrics, expand the output.
RTA Primary Analysis Metrics Captured by MiSeq Run (MiSeq v3.2) Step
Per Read LIMS Field Name (Stored on derived sample/analyte input to the step) | Per Lane LIMS Field Name (Stored in measurement placeholders in the Sample Details table on the Record Details screen) |
% Aligned R1 | % Aligned R1 |
% Aligned R2 | % Aligned R2 |
% Bases >=Q30 R1 | % Bases >=Q30 R1 |
% Bases >=Q30 R2 | % Bases >=Q30 R2 |
% Error Rate R1 | % Error Rate R1 |
% Error Rate R2 | % Error Rate R2 |
% Phasing R1 | |
% Phasing R2 | |
% Prephasing R1 | |
% Prephasing R2 | |
%PF R1 | |
%PF R2 | |
Cluster Density (K/mm^2) R1 | |
Cluster Density (K/mm^2) R2 | |
Intensity Cycle 1 R1 | Intensity Cycle 1 R1 |
Intensity Cycle 1 R2 | Intensity Cycle 1 R2 |
Reads PF (M) R1 | |
Reads PF (M) R2 | |
Yield PF (Gb) R1 | Yield PF (Gb) R1 |
Yield PF (Gb) R2 | Yield PF (Gb) R2 |
How It Works
The sequencing service runs on the Clarity LIMS server. The service detects event files that instrument RTA produces as the run progresses. The event files let the service know where to find the run data.
As the run data are written out and the End Run event is detected, the following events occur:
The data are matched to the step based on the reagent cartridge ID that was entered or scanned on the Denature and Dilute (MiSeq v3.2) step.
Read-only field values on the Record Details screen are populated accordingly.
When the service has finished processing the end run event and updating the fields in Clarity LIMS, the sequencing service generates the report and attaches it to the step.
Scripts and Files Installed
This integration requires components installed with the Illumina Preset Protocols (IPP).
The Illumina MiSeq Integration Package v8.3.0 RPM installs the scripts and files listed in the following table.
Component | Location | Description |
configure_extensions_miseq_sequencingservice-v8.sh | /opt/gls/clarity/config/ | Script that installs the service properties in the database. |
log4j.xml | /opt/gls/clarity/extensions/Illumina_MiSeq/v8/SequencingService/conf | File containing the settings for logging the sequencing jar. |
miseq-sequencing-v8.jar | /opt/gls/clarity/extensions/Illumina_MiSeq/v8/SequencingService | Jar file containing API-based Clarity LIMS extensions used for capturing run results and report generation. |
InterOp libraries | /opt/gls/clarity/extensions/Illumina_MiSeq/v8/lib | Illumina shared library for parsing InterOp data files. |
MiSeqSamplesheetv3.csv | /opt/gls/clarity/extensions/conf/driverfiletemplates | Template file used for file generation. |
Properties Installed
Refer to Integration Properties Details for the properties installed with Illumina MiSeq Integration v8.3.0.
Consumables Installed
Reagent Categories and Kits
Reagent categories or label groups are installed with the IPP workflow configuration slices.
The MiSeq Reagent Kit is included in the Illumina MiSeq Integration.
Control Types
The PhiX v3 control type is included in the Illumina MiSeq Integration.
Container Types
The following container types are included in the Illumina MiSeq Integration:
MiSeq Reagent Cartridge
96 well plate
Tube
All one-dimensional container types with both numeric rows and numeric columns are supported.
Instrument Integration
To make sure that the Illumina instrument warranty remains valid, the instrument integration must be performed and maintained by the Clarity LIMS Support team. To perform this integration, the Support team requires remote access to the system while it is idle.
The following steps are performed by the Clarity LIMS Support team when configuring the sequencing for use with the Illumina MiSeq Integration.
Create a directory on the local computer to hold the batch files. These batch files write event files to the network attached storage (NAS) shares.
Create a directory on the NAS to hold the event files.
Modify Illumina software configuration files to call the batch files that create the event files.
Update sequencing service default properties to match the specifics of the installation.
Rules and Constraints
The Illumina MiSeq Integration operates with the following constraints:
The reagent cartridge ID must be unique. There should not be multiple reagent cartridge containers in the system with the same name.
The reagent cartridge ID must be scanned as the reagent cartridge Container Name on the Denature & Dilute (MiSeq v3.2) step.
Last updated